TP112 Models:
Model | Input | Output |
TP112-N | TTL | TTL |
TP112-R | optical isolated | relay (5 A / 50 V) |
TP112-S | optical isolated | solid state (5 A / 50 V) |
Specifications
The tables below show detailed specifications of the TP112.
Acquisition system (analog) | |||
---|---|---|---|
Number of input channels | 2 analog, D-Sub | ||
Resolution | 12 bit @ 1 MHz | ||
Accuracy | 2% of full scale ± 1 LSB | ||
Ranges | 1.25 V to 20 V full scale, each channel | ||
Coupling | AC/DC | ||
Impedance | 1 MOhm / 20 pF | ||
Protection (in all ranges) | 200 V AC 50 Hz (1:1 probe) 600 V AC 50 Hz (1:10 probe) | ||
Bandwidth (-3dB) | DC to 500 kHz maximum | ||
AC coupling cut off frequency (-3dB) | ±1.5 Hz | ||
Maximum sampling rate | 1 MHz, 1 µs | ||
Sampling clock source | internal, ±0.01% | ||
Memory | 32768 samples each channel | ||
Acquisition system (digital) | |||
model TP112-N | model TP112-R | model TP112-S | |
Input channels | 8 digital, D-Sub | ||
Voltage "low" | 0 to 0.8 V | 0 to 1 V | 0 to 1 V |
Voltage "high" | 2.4 to 5 V | 3 to 24 V | 3 to 24 V |
Impedance | 100 kOhm | 1 kOhm | 1 kOhm |
Bandwidth | 1 MHz | 5 kHz | 5 kHz |
Outputs (digital) | |||
model TP112-N | model TP112-R | model TP112-S | |
Output channels | 8 digital, D-Sub | ||
Voltage | TTL | 50 V max | 50 V max |
Current | 0 to 5 mA | 0 to 5 A | 0 to 5 A |
Impedance | 100 kOhm | 1 Ohm | 1 Ohm |
Bandwidth | 1 MHz | 10 Hz | 1 kHz |
Trigger system | |||
System | digital, 2 levels | ||
Source | CH1, CH2, digital input | ||
Trigger modes | rising edge, falling edge | ||
Level adjustment | 0 to 100% of full scale | ||
Hysteresis adjustment | 0 to 100% of full scale | ||
Resolution | 0.39% (8 bits) | ||
Pre trigger | 0 to 32768 samples (0 to 100%) | ||
Post trigger | 0 to 32768 samples (0 to 100%) | ||
Interface | |||
Interface | 8 bit ISA | ||
Power Requirements | |||
Power from ISA slot | 7.6 W max +5 V 800 mA +12 V 200 mA -12 V 100 mA | ||
Physical | |||
Instrument height | 125 mm (4.9 inch) | ||
Instrument length | 250 mm (9.8 inch) | ||
Instrument width | 20 mm (0.8 inch) | ||
Weight | 230 g (8.1 ounce) | ||
Accessories | 2 oscilloscope probes 1:1-1:10 switchable D-Sub to BNC adapter |
Downloads
Download software, drivers, manuals and source code for the TP112.
Software & Driver
Description | Version | Remarks | Size | |
---|---|---|---|---|
LPT/ISA Driver (GiveIO.sys) | Windows NT4 / 2000 / XP, 32 bit | 247.57 kB | download | |
TP112 WinSoft measurement software (32 bit) | 2.87 | Windows 95/98/ME/NT4/2000/XP | 1.94 MB | download |
TP112 WinSoft measurement software (16 bit) | 2.30 | Windows 3.xx | 2.14 MB | download |
TP112 DOS Measurement software | 1.0 | DOS protected mode, Dutch | 144.34 kB | download |
TP112 DOS Measurement software | 2.0 | DOS protected mode, English | 254.97 kB | download |
Manuals
Programming
Description | Version | Remarks | Size | |
---|---|---|---|---|
TP112 Instrument DLL (32 bit) | 2.90 | Windows 95/98/ME/NT4/2000/XP | 448.67 kB | download |
TP112 Instrument DLL (16 bit) | Windows 3.xx | 97.00 kB | download | |
Borland C++ Builder example | 18.53 kB | download | ||
Borland Delphi example | 15.38 kB | download | ||
DOS source code archive | Basic / C / Pascal | 179.18 kB | download | |
DasyLab driver | For all TiePie measuring instruments | 259.46 kB | download | |
Instrument DLL programmer's manual | 1.29 | English | 331.35 kB | download |
Matlab example | for Matlab 6.5.1 and higher | 2.80 kB | download | |
Python binding for instrument DLL | 1.7 | for Python 2.7 (Windows 32 bit) | 200.51 kB | download |
Python examples | 1.7 | 4.90 kB | download | |
Visual Basic 6 example | 12.84 kB | download |